Field laboratory comparison of STANDARD Q Filariasis Antigen Test (QFAT) with Bioline Filariasis Test Strip (FTS) for the detection of Lymphatic Filariasis in Samoa, 2023

0

by Jessica L. Scott, Helen J. Mayfield, Jane E. Sinclair, Beatris Mario Martin, Maddison Howlett, Ramona Muttucumaru, Kimberly Y. Won, Robert Thomsen, Satupaitea Viali, Rossana Tofaeono-Pifeleti, Patricia M. Graves, Colleen L. Lau

Background

To monitor the progress of lymphatic filariasis (LF) elimination programmes, field surveys to assess filarial antigen (Ag) prevalence require access to reliable, user-friendly rapid diagnostic tests. We aimed to evaluate the performance of the new Q Filariasis Antigen Test (QFAT) with the currently recommended Filariasis Test Strip (FTS) for detecting the Ag of Wuchereria bancrofti, the causative agent of LF, under field laboratory conditions.

Methodology/Principal findings

During an LF survey in Samoa, 344 finger-prick blood samples were tested using FTS and QFAT. Microfilariae (Mf) status was determined from blood slides prepared from any sample that reported Ag-positive by either Ag-test. Each test was re-read at 1 hour and the next day to determine the stability of results over time. Overall Ag-positivity by FTS was 29.0% and 30.2% by QFAT. Concordance between the two tests was 93.6% (kappa = 0.85). Of the 101 Mf slides available, 39.6% were Mf-positive, and all were Ag-positive by both tests. Darker test line intensities from Ag-positive FTS were found to predict Mf-positivity (compared to same/lighter line intensities). QFAT had significantly higher reported test result changes than FTS, mostly reported the next day, but fewer changes were reported between 10 minutes to 1-hour. The field laboratory team preferred QFAT over FTS due to the smaller blood volume required, better usability, and easier readability.

Conclusion/Significance

QFAT could be a suitable and user-friendly diagnostic alternative for use in the monitoring and surveillance of LF in field surveys based on its similar performance to FTS under field laboratory conditions.